Chapter 4: Operation
62
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
4.8.2 Turning AquaA EMERGENCY MODE on
The EMERGENCY MODE is initiated by the following steps:
Step 1
Turn the reverse osmosis system and all other devices (
AquaA2
,
AquaHT
etc.) off using the main power switch (
OFF/O
position)
Step 2
Change the position of the emergency operation switch
The button can be turned either to the left or to the right.
This selects the power supply unit and turns pump
P1
on.
Step 3
Return the main power switch of the
AquaA
to the
ON/I
position.
Information on booting the system (see chapter 4.3.1 on page 44).
Step 4
This step deactivates the actuator outputs of the electronic control so
that no actions can be performed.
If pump
P1
did not turn on, set the switch to the other position. To do
so, restart at step 1 and use the emergency operation switch setting
2 in step 2.
Step 5
By pressing the button for pump
P2
, the produced volume of dialysis
water can be increased.
Power supply unit 1
Power supply unit 2
Normal position
Summary of Contents for AquaA
Page 2: ......
Page 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Page 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...