Chapter 5: Alarms
128
Fresenius Medical Care
AquaA
IFU-EN 06A-2021
F–02–02–02
Signal:
red
FAILURE Stage 2,
permeate temp.
alarm limit
exceeded
– The dialysis water
temperature has exceeded
the specified alarm limit.
– Conductivity sensor CDT-Ps
defective
– Measuring transducer B4
defective
Contact the service
department
This failure is
confirmed
automatically when
the value drops
below the alarm
limit. The message
will continue to be
shown on the
display, however.
F–02–02–03
Signal:
red
FAILURE Stage 2,
permeate pressure
alarm limit
exceeded
– The dialysis water pressure
has exceeded the specified
alarm limit.
– Pressure sensor P-Ps
defective
– Measuring transducer B4
defective
Contact the service
department
F–02–02–04
Signal:
red
FAILURE Stage 2,
conc. pressure
alarm limit
exceeded
– The concentrate pressure
has exceeded the specified
alarm limit.
– Pressure sensor P-Cs
defective
– Measuring transducer B4
defective
Contact the service
department
F–02–02–05
Signal:
red
FAILURE Stage 2,
run-dry protection,
pump stop
– The
AquaA2
pre-pressure
P-Fs is below the specified
alarm limit.
–
AquaA
not producing
enough dialysis water
–
AquaA
membranes
defective
Contact the service
department
F–02–02–08
Signal:
red
FAILURE Stage 2,
leakage detector
signals leakage
– Indication of water leakage
from the connected leakage
sensor
Check all water-
carrying lines and
connections.
Contact the service
department
W–02–02–01
Signal:
yellow
WARNING Stage 2,
Permeate cond.
alarm limit
exceeded
– The dialysis water
conductivity CD-Ps has
exceeded the specified
alarm limit.
– Conductivity sensor CD-Ps
defective
– Measuring transducer B4
defective
This failure is
confirmed
automatically when
the value drops
below the alarm
limit.
Contact the service
department
Error code
Visual
indica-
tor
Messages
Cause
Action required
Summary of Contents for AquaA
Page 2: ......
Page 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Page 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...