Chapter 5: Alarms
126
Fresenius Medical Care
AquaA
IFU-EN 06A-2021
W–03–04–02
Signal:
yellow
WARNING:
Start, pressure
cannot be built up
– The concentrate pressure
failed to rise above the
specified alarm limit.
– Pressure sensor P-C
defective
– Booster pump P1 defective
– The motor protection switch
has tripped.
Contact the service
department
W–03–04–03
Signal:
yellow
WARNING:
Start, no circulation
flow
– The circulation pump P3
failed to generate a flow.
– Flow control switch P3ctrl
defective
Contact the service
department
W–03–04–04
Signal:
yellow
WARNING:
Start, permeate
cond. too high
– The dialysis water
conductivity CD-P failed to
drop below the specified
alarm limit.
– Conductivity sensor CD-P
defective
– Measuring transducer B4
defective
Contact the service
department
Error code
Visual
indica-
tor
Messages
Cause
Action required
Summary of Contents for AquaA
Page 2: ......
Page 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Page 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...