Chapter 12: Specifications
168
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
Water pretreatment
Pretreatment of the water is determined after previous water analysis.
Sensors
Parameter
Values
Unit
Water hardness
< 1 °dH
Total chlorine
< 0.1
mg/L
Iron*
< 0.1
mg/L
Manganese*
< 0.05
mg/L
Silicate*
< 25
mg/L
Max. conductivity
2500
µS/cm
SDI*
(
S
ilt-
D
ensity
I
ndex or colloid index)
< 3
- - -
pH
6 to 8
- - -
* The parameters for iron, manganese, silicate, and SDI in the inlet water
should be checked before the water pretreatment components are
dimensioned.
Warning
Risk for the patient due to damage to the membrane
Deviating water quality can reduce the service life of the membrane.
This may require the membrane to be replaced prematurely.
Ensure compliance with the necessary parameters.
Measured value
Sensor
Measuring
range
Unit
Accuracy
Dialysis water conductivity
CD–P
0.0 to 100.0
µS/cm
±5 % of the MV*;
±0.1 µS/cm
100 to 2500
±10 % of the
MV*; ±0.1 µS/cm
Inlet conductivity
CD–F
0.0 to 100.0
µS/cm
±5 % of the MV*;
±0.1 µS/cm
100 to 2500
±10 % of the
MV*; ±0.1 µS/cm
Dialysis water temperature
T–P
0.0 to 115.0
°C
±2 °C
Inlet temperature
T–F
0.0 to 115.0
°C
±2 °C
Dialysis water pressure
P–P
0.0 to 10.0
bar
±1%
Concentrate pressure
P–C
0.0 to 20.0
bar
±1%
Summary of Contents for AquaA
Page 2: ......
Page 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Page 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...