Chapter 15: Appendix
226
Fresenius Medical Care
AquaA
IFU-EN-US 06B-2021
Operating modes and functions
A
– Select program (
STANDBY
,
SUPPLY
,
RINSE
)
– Press and hold the button for 3 seconds
IFU
B
Operating programs (see
–
STANDBY
–
SUPPLY
–
RINSE
(Cleaning as well as the sub-operating modes Rinse
AquaA
and Rinse water
pretreatment system)
–
EMERGENCY MODE
IFU
C
Status menu
(no password required) (see
– Messages:
Current messages
Confirm messages
– Report
– Start, Stop:
Switching program settings for Supply and Rinse
Changing the autostop time once
– System information:
Configuration and system values
– Operating data (current operating data)
IFU
D
Emergency mode (see
– No emergency supply with soft water
– Dialysis water conductivity and temperature in the break tank are monitored
– Activating the emergency mode:
Turn the reverse osmosis system and any options off using the main power switch
(OFF)
Turn the
AquaA
emergency mode switch counter-clockwise or clockwise
Turn the
AquaA
reverse osmosis system on using the main power switch (ON)
The second pump can be switched on by pressing this switch–even if pump 1 fails to
switch on.
– Deactivating the emergency mode:
Set the main power switch of the
AquaA
to the OFF position
Set the emergency switch to its center position
Return the main power switch of the
AquaA
reverse osmosis system and options
to
the ON position
IFU
E
AquaA2
emergency mode
– No emergency supply with soft water
– Dialysis water conductivity is monitored
– Activating the emergency mode:
–
Turn the
AquaA
reverse osmosis system and options off using the main power
switch (OFF)
–
Change the water supply path to Emergency mode 2
–
Turn the
AquaA2
emergency mode switch clockwise
–
Turn the
AquaA2
reverse osmosis system on using the main power switch (ON)
–
The second pump can be switched on by pressing this switch–even if pump P1s
fails to switch on
– Deactivating the emergency mode:
–
Turn the
AquaA2
reverse osmosis system off using the main power switch
(OFF)
–
Change the water supply path to Standard operation 1
–
Set the emergency switch to its center position
–
Return the main power switch of the
AquaA
,
AquaA2
and options to the ON
position. Before restoring dialysis operation, a disinfection (chemical or
membrane heat disinfection) must be performed.
IFU
Alarms
A
Visual indicator (see
– Red and flashing–an alarm or a malfunction has occurred and has not yet been
confirmed
– Yellow and flashing–a warning has occurred and has not yet been confirmed
– Yellow–Disinfection or service in progress
– Green–
SUPPLY
mode is active
– Green and flashing–the system is in the process of switching to the
SUPPLY
mode
IFU
B
Error messages (see
– Are displayed directly if an alarm occurs
– Refer to Chapter 5 "Alarms" in the Instructions for Use
– Refer to Chapter 2 "Addresses" (Water Technology hotline)
IFU
Training subjects
Filed in
Summary of Contents for AquaA
Page 2: ......
Page 10: ...10 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 32: ...Chapter 2 Important information 32 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 108: ...Chapter 4 Operation 108 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 152: ...Chapter 9 Installation 152 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 186: ...Chapter 13 Definitions 186 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...
Page 217: ...Chapter 14 Options Fresenius Medical Care AquaA IFU EN US 06B 2021 217 ...
Page 218: ...Chapter 14 Options 218 Fresenius Medical Care AquaA IFU EN US 06B 2021 ...