48
Manual # 26-0111000-00 / Revision D
ScreenPro PLUS – Multi-Screen Seamless Switcher
SYSTEM
Æ
RECALL OPTIONS
Recall options allow the user to select which items are recalled from Preset Memories when a system preset key is
pressed. For example, if the Recall Option Effect is turned OFF, the Transition Effect will not be recalled from Preset
Memories. Use the +/ON and -/OFF
keys to configure the option.
SYSTEM
Æ
DIAGNOSTICS
This menu is used to test the controller’s hardware. Use the UP and DOWN arrow keys to highlight the desired test
to execute and press the RIGHT arrow key.
SYSTEM
Æ
DIAGNOSTICS
Æ
VFD TEST
This menu is used to test the controller’s VFD display hardware. Use the UP and DOWN arrow keys to highlight the
desired test to execute and press the +/ON key.
SYSTEM
Æ
DIAGNOSTICS
Æ
VFD TEST
Æ
BRIGHTNESS TEST
This diagnostic test verifies the VFD display hardware by changing the display brightness through all
possible settings. The display should change from dim to bright in 16 steps. The information on the
screen should not flicker, drop out, or change other than in intensity. If the display shows traces of
previous display information, the brightness setting is probably too high and should be reduced to avoid
further screen burn-in.
Å
RECALL OPTIONS
PREVIEW ON
EFFECT ON
RATE ON
AUX ON
Å
DIAGNOSTICS
VFD TEST
Æ
JOYSTICK / TBAR TEST
Æ
KEY DETECTION TEST
Æ
LED TEST
Æ
Å
VFD TESTS
BRIGHTNESS TEST [+]
PIXEL TEST [+]