NetDAQ
Service Manual
1-20
Table 1-27. 2640A Frequency Sensitivity Specifications
Frequency Measurement Sensitivity (Sinewave)
Frequency Range
Minimum Signal
Maximum Signal
15 Hz to 70 kHz
100 mV ac rms
V<150/300Vrms [1] and Vx Hz<2x10
6
)
70 kHz to 100 kHz
100 mV ac rms
20V ac rms
100 kHz to 200 kHz
150 mV ac rms
10V ac rms
200 kHz to 300 kHz
150 mV ac rms
7V ac rms
300 kHz to 1 MHz
Linearly increasing from 150 mV ac
rms at 300 kHz to 2 V ac rms at 1 MHz
Linearly decreasing from 7 V ac rms
at 300 kHz to 2 V ac rms at 1 MHz
[1] 300V range applies to channels 1 and 11 only.
2645A Specifications
1-29.
This section includes specifications specific to the 2645A instrument by measurement
function.
2645A DC Voltage Measurement Specifications
1-30.
Tables 1-28 to 1-30 provide 2645A specifications for the dc voltage measurement
function.
Table 1-28. 2645A DC Voltage General Specifications
Specification
Characteristic
Input Impedance
100 M
Ω
in parallel with 300 pF maximum for ranges <=3V
10 M
Ω
in parallel with 100 pF maximum for ranges >3V
Normal Mode Rejection
50 dB minimum at 50 Hz/60 Hz +0.1%, Slow Rate
Common Mode Rejection
120 dB minimum at dc, 50 Hz/60 Hz +0.1%, 1 k
Ω
imbalance, Slow Rate
80 dB minimum at dc, 60 dB at 50 Hz/60 Hz +0.1%, 1 k
Ω
imbalance,
Medium and Fast Rates
Channel-to-Channel
Crosstalk
120 dB minimum Slow Rate (e.g., 30V dc on channel 1 may cause a 30
µ
V
error on channel 2)
80 dB minimum Medium and Fast Rates (e.g., 1V dc on channel 1 may
cause a 10
µ
V error on channel 2)
Temperature Coefficient
For % input: Add 1/10th the 90-day specification per
°
C above 28
°
C or
below 18
°
C.
For floor error (V): Add 1/20th the 90-day specification per
°
C above 28
°
C
or below 18
°
C.
Maximum Input Voltage
The lesser voltage of:
50V dc or 30V ac rms from any input terminal to earth
-or-
50V dc or 30V ac rms from any input terminal to any other input terminal
Summary of Contents for NetDAQ 2640A
Page 14: ...NetDAQ Service Manual x...
Page 46: ...NetDAQ Service Manual 2 4...
Page 106: ...NetDAQ Service Manual 2 64...
Page 108: ...NetDAQ Service Manual 3 2...
Page 164: ...NetDAQ Service Manual 4 42...
Page 206: ...NetDAQ Service Manual 6 2...
Page 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Page 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Page 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Page 234: ...NetDAQ Service Manual 7 2...
Page 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Page 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Page 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Page 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Page 260: ...2640A 2645A Service Manual 7 28...