General Maintenance
Line Fuse
3
3-5
Replacing the Line Fuse
3-9.
The line fuse (15/100 ampere, 250V, time delay, PN 944629) is in series with the power
supply and located inside the instrument. To replace the fuse, refer to Figure 3-1 and the
following procedure:
WARNING
Do not operate the instrument without the cover properly
installed.
1.
Disconnect all rear panel cables to the instrument power, Universal Input Module,
and I/O connectors.
2.
Invert the instrument on a protective surface and remove the four 1/4-inch 6-32
Phillips-head screws on the bottom of the case.
3.
Turn the instrument upright and remove the two 1/2-inch 6-32 Phillips-head screws
from the rear panel bezel.
4.
Remove the rear panel bezel and case assembly. Do not touch any internal parts of
the instrument!
5.
Locate the fuse holder at the back of the chassis near the power input connector.
Using a non-metallic tool, carefully pry the fuse from the holder.
6.
Insert the new fuse into the holder. (You must use a 15/100 ampere, 250V, time
delay line fuse replacement, PN 944629.)
7.
Reinstall the case to its original position (the rubber feet are towards the front of the
instrument).
8.
Reinstall the rear panel bezel (rubber feet towards the bottom) and attach it with the
two 1/2-inch 6-32 Phillips-head screws.
9.
Invert the instrument on the protective surface and reinstall the four 1/4-inch 6-32
screws on the bottom securing the case.
10.
Reinstall the cables removed in Step 1.
Summary of Contents for NetDAQ 2640A
Page 14: ...NetDAQ Service Manual x...
Page 46: ...NetDAQ Service Manual 2 4...
Page 106: ...NetDAQ Service Manual 2 64...
Page 108: ...NetDAQ Service Manual 3 2...
Page 164: ...NetDAQ Service Manual 4 42...
Page 206: ...NetDAQ Service Manual 6 2...
Page 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Page 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Page 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Page 234: ...NetDAQ Service Manual 7 2...
Page 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Page 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Page 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Page 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Page 260: ...2640A 2645A Service Manual 7 28...