![Fluke NetDAQ 2640A Service Manual Download Page 176](http://html1.mh-extra.com/html/fluke/netdaq-2640a/netdaq-2640a_service-manual_2310217176.webp)
NetDAQ
Service Manual
5-12
Command Set
5-18.
The instrument RS-232 command set is shown in Table 5-7: RS-232 Command Set. The
calibration commands are described in more detail in Chapter 4. These are: CAL,
CAL_CLR, CAL_CONST, CAL_CONST?, CAL_REF, CAL_REF?, and CAL_STEP?.
The other commands are described in this chapter. The instrument does not accept the
calibration commands unless calibration mode is enabled. In calibration mode, the
instrument does not accept some non-calibration commands. It is expected that users will
make all calibration adjustments and exit calibration mode before performing calibration
verification. Calibration verification readings will be made in the slow reading rate,
which is the power-on and reset default, therefore no command is provided for setting
the rate.
Table 5-7. RS-232 Command Set
Command
Description
Cal
Not Cal
*IDN?
Identification query
*
*
*OPC?
Operation complete query
*
*
*RST
Reset
*
*
*TST?
Self test query
*
SELFTEST?
Return current selftest results
*
*
CAL
Start calibration procedure for indicated
function
*
CAL?
Return the identifier of any CAL procedure in
progress
*
CAL_CLR
Reset calibration constants to nominal value
*
CAL_CONST?
Query the value of a particular calibration
constant
*
*
CAL_REF
Specify value to calibrate to (in place of default
reference value)
*
CAL_REF?
Query the present calibration reference value
*
CAL_STEP?
Calibrate and query the calibrated value of the
input
*
FUNC
Configure function, range, terminals for channel
1
*
FUNC?
Query function, range, terminals for channel 1
*
MEAS?
Trigger and query a measurement on channel 1
*
The RS-232 commands are described in this chapter using the following format:
•
Command: The command name and syntax.
•
Description: A description of the command.
•
Parameters: A description of the required and optional parameters.
•
Response: A description of data returned.
•
Restrictions: When the command is allowed/disallowed.
•
Notes: Any additional useful information.
Summary of Contents for NetDAQ 2640A
Page 14: ...NetDAQ Service Manual x...
Page 46: ...NetDAQ Service Manual 2 4...
Page 106: ...NetDAQ Service Manual 2 64...
Page 108: ...NetDAQ Service Manual 3 2...
Page 164: ...NetDAQ Service Manual 4 42...
Page 206: ...NetDAQ Service Manual 6 2...
Page 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Page 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Page 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Page 234: ...NetDAQ Service Manual 7 2...
Page 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Page 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Page 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Page 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Page 260: ...2640A 2645A Service Manual 7 28...