NetDAQ
Service Manual
3-16
Assembling the Front Panel Assembly
3-31.
Complete the following procedure to assemble the Front Panel Assembly [D].
1.
Place the Front Panel Assembly on a protective surface to prevent any scratching or
damage to the assembly.
2.
Position the display window [H] in place; then push it firmly to snap it into position
in the assembly.
3.
Position the elastomeric keypad [G] in place; then gently press it into position in the
assembly.
4.
Position the A2 Display PCA [F] in place; then push it firmly to snap it into position
in the assembly. It may be necessary to slightly distend the top of the bezel to allow
the pca to snap into position.
NOTE
The Display PCA provides space for securing screws, which are
used only if one or more tabs are broken. If the pca uses one or
more securing screws, install these as well.
CAUTION
Avoid using ammonia or methyl-alcohol cleaning agents
on either the Front Panel or the display window. These
types of cleaners can damage surface features and
markings. Use an isopropyl-based cleaning agent or
water to clean the Front Panel and the display window.
Installing the Front Panel Assembly
3-32.
Complete the following procedure to install the Front Panel Assembly [D]:
1.
Gently connect the display ribbon cable [E] to the connector on A2 Display PCA.
2.
Position the Front Panel Assembly on the chassis mounting post hardware [Y].
3.
Complete the procedure “Installing the Instrument Case,” as required.
Installing the Instrument Case
3-33.
Complete the following procedure to install the instrument case [A]. (Refer to
Figure 3-2.)
WARNING
Do not operate the instrument without the cover properly installed.
1.
Reinstall the case to its original position (the rubber feet are towards the front of the
instrument).
2.
Reinstall the rear panel bezel (rubber feet towards the bottom) [Z] and attach it with
the two 1/2-inch 6-32 Phillips-head screws [C].
3.
Invert the instrument on the protective surface and reinstall the four 1/4-inch 6-32
screws [B] on the bottom securing the case.
4.
Reinstall the measurement connection cables and power cord, as required.
Summary of Contents for NetDAQ 2640A
Page 14: ...NetDAQ Service Manual x...
Page 46: ...NetDAQ Service Manual 2 4...
Page 106: ...NetDAQ Service Manual 2 64...
Page 108: ...NetDAQ Service Manual 3 2...
Page 164: ...NetDAQ Service Manual 4 42...
Page 206: ...NetDAQ Service Manual 6 2...
Page 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Page 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Page 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Page 234: ...NetDAQ Service Manual 7 2...
Page 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Page 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Page 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Page 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Page 260: ...2640A 2645A Service Manual 7 28...