Performance Testing and Calibration
Performance Test
4
4-23
Totalizer Sensitivity Test
4-26.
This test checks the ability of the Totalizer feature to count voltage transition at a
particular sensitivity level.
1.
Connect Test Leads At the DIGITAL I/O connector, connect the
Σ
(Totalizer) test
lead and GND test lead to a signal generator’s output terminals. Adjust the signal
generator for an output of 1.5V rms sine wave at 10 Hz.
2.
Start Instrument Scanning Click the Start Instrument button on the Button Bar to
start instrument scanning. Scanning is initiated to enable the return of TOTAL status
to the Spy window.
3.
Open Spy Window Select the Spy command from the Utilities menu. Select
01TOTAL. Click OK to open the Spy window.
4.
Verify Totalizer Count The Spy window summarizes the Totalizer count. Verify
the totalizer count is advancing at approximately 10 Hz per Spy window update
(nominal 1 second intervals).
5.
Close Spy Window To close the Spy window, double-click the upper left-hand
corner control-menu box.
6.
Stop Scanning Click the Stop Instrument button on the Button Bar to stop
instrument scanning.
Master Alarm Output Test
4-27.
This test checks the Master Alarm output for a logic low when a channel is in alarm.
1.
Connect Test Leads to ALARM/TRIGGER I/O Connector Remove the
8-position ALARM/TRIGGER I/O connector from the instrument rear panel.
Connect a test lead to each line, MA (Master Alarm), TO (Trigger Output), TI
(Trigger Input), plus a test lead to the GND line. Reinstall the connector.
2.
Measure Unset MA Line Using a digital multimeter, measure the output of the
unset MA test lead, referenced to the GND test lead, for a voltage greater than +3.8V
dc.
3.
Verify Configuration Channel 1 for Volts DC In NetDAQ Logger for Windows,
verify channel 1 is configured for Volts dc, 3V range.
4.
Verify Configuration Channel 1 for Alarms In NetDAQ Logger for Windows,
verify channel 1 is configured for an Alarm 1 with Alarm Sense=LO, Alarm
Value=1 and Digital Output=DO0.
5.
Start Instrument Scanning Click the Start Instrument button on the Button Bar to
start instrument scanning. Scanning is initiated to enable the Master Alarm output.
6.
Measure Set MA Line Using a digital multimeter, measure the output of the set
MA test lead, referenced to the GND test lead, for a voltage less than +0.8V dc.
7.
Stop Scanning Click the Stop Instrument button on the Button Bar to stop
instrument scanning.
Summary of Contents for NetDAQ 2640A
Page 14: ...NetDAQ Service Manual x...
Page 46: ...NetDAQ Service Manual 2 4...
Page 106: ...NetDAQ Service Manual 2 64...
Page 108: ...NetDAQ Service Manual 3 2...
Page 164: ...NetDAQ Service Manual 4 42...
Page 206: ...NetDAQ Service Manual 6 2...
Page 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Page 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Page 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Page 234: ...NetDAQ Service Manual 7 2...
Page 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Page 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Page 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Page 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Page 260: ...2640A 2645A Service Manual 7 28...