3. OPERATION
Page 71
© 1998-2007 DH Instruments, a Fluke Company
Table 13.
molbloc-L Size and Nominal Range Designations
molbloc
“NOMINAL
RANGE”
DESIGNATION
molbloc-L SIZE
DESIGNATION
10 sccm
1E1
50 sccm
5E1
100 sccm
1E2
200 sccm
2E2
500 sccm
5E2
1 slm
1E3
5 slm
5E3
10 slm
1E4
30 slm
3E4
100 slm
1E5
molbloc-S elements are also identified by size designations, each of which relate
to a specific molbloc-S K
F
value. molbloc-S flow ranges depend on calibration
type and the pressure limitations of the application and molbox used. For
information on the possible molbloc-S flow ranges with various operating
pressures, see Section 1.2.4.2.2). molbloc-S size and K
F
value correspondence are
given in Table 14.
Table 14.
molbloc-S Size Designation and Pressure to Flow Conversion Ratio (K
F)
K
F
(sccm/kPa)
molbloc-S SIZE
DESIGNATION
50
5E1-S
100
1E2-S
200
2E2-S
500
5E2-S
1 000
1E3-S
2 000
2E3-S
5 000
5E3-S
10 000
1E4-S
3.4.9
[RES]
PURPOSE
To set the resolution of molbox RFM’s display of the flow through the molbloc and other flow
display and entry values.
PRINCIPLE
The resolution with which the flow measured by molbox RFM is displayed can be adjusted. This
feature can be used to reduce the resolution when lower precision measurements are being
made and additional digits might confuse or distract the operator.
The resolution setting determines the number of digits with which flow is displayed. The desired
resolution is calculated based on the nominal nitrogen gas full scale of the molbloc range in
the current flow unit of measure and then rounded to the furthest digit to the right (i.e., resolution
of 0.01 % on a 100 sccm molbloc is 0.01 sccm).