3. OPERATION
Page 81
© 1998-2007 DH Instruments, a Fluke Company
The High security level disables remote communications and returns an error
message (“ERROR”) to all remote commands. All other security levels have
NO effect on remote communications.
The security levels are structured to support typical levels of operation.
Specifically, the security levels prevent execution of the functions accessed by
the key strokes marked by “
•
”.
Table 16.
Security Levels - Functions NOT Executed Per Function/Level
KEYS
LOW
MEDIUM
HIGH
[K]
•
[GAS]
•
[UNIT]
•
[UNIT]
(change temperature/pressure conditions)
•
•
[TARE]
(access menu)
•
[TARE], <1tare>, <select tare pressure>
•
•
[TARE], <2purge>,
(change purge time)
•
•
[TARE], <4AutoZ>, <1on/off>
•
•
•
[TARE], <4AutoZ>, <3edit>
•
•
•
[TARE], <4AutoZ>, <4run>
•
•
[P&T]
•
[DISPLAY]
•
[DISPLAY],
(change times/target)
•
•
[MICRO]
•
•
•
[molbloc]
•
[RES]
•
•
[SETUP], <1flowU>
•
•
[SETUP], <2presU>
•
•
[SETUP], <3tempU>
•
•
[SETUP], <4molbloc>
•
[SETUP], <5stab>
•
•
[SETUP], <6adj>
•
•
•
[SPECIAL], <1reset>
•
•
[SPECIAL], <1reset>, <3com>
•
•
•
[SPECIAL], <1reset>, <4cal>
•
•
•
[SPECIAL], <1reset>, <5all>
•
•
•
[SPECIAL], <4cal>
•
•
[SPECIAL], <4cal>, <any 3edit>
•
•
•
[SPECIAL], <5prefs>
•
[SPECIAL], <5prefs>, <1ScrSvr>
•
•
[SPECIAL], <5prefs>, <2sound>
•
•