4. REMOTE OPERATION
Page 111
© 1998-2007 DH Instruments, a Fluke Company
FUNIT(=)
Purpose
Read or set the unit of measurement for the flow through the
molbloc as calculated by the molbox RFM.
Syntax “FUNIT=
unit
”
“FUNIT”
Arguments “FUNIT=sccm”
Defaults
unit:
The flow unit used to display the measured flow.
Remarks
The flow unit displayed can be changed using this command.
The flow unit must be supported by the molbox RFM. The flow
unit protocol is the same as the front panel flow unit definitions.
Example Command:
Reply:
“FUNIT=SLM”
“SLM”
Errors
ERR# 7:
The flow unit is invalid or not supported by the
molbox RFM.
See Also
3.2.6.5, “UTEMP”, “VLM”
GAS(=)
Purpose
Read or set the gas type being flowed through the molbloc.
Syntax “GAS=
gas
”
“GAS”
Arguments
gas:
The gas to use.
Remarks
The gas type being flowed through the molbloc is specified
using this command. The gas must be supported by the molbox
RFM. The gas type protocol is the same as the front panel gas
definitions. A valid molbloc must be connected to the molbox
RFM before changing the gas type.
Example Command:
Reply:
“GAS=AR”
“Ar”
Errors ERR#
7:
ERR# 24:
ERR# 32:
ERR# 35:
ERR# 39:
ERR# 40:
ERR# 41:
The gas is invalid for the molbox RFM.
A molbloc is not connected to the molbox RFM.
This gas is not supported by the molbox RFM.
The molbloc experienced a communications time
out.
The molbloc experienced a data verification
failure.
The molbloc experienced a read checksum
failure.
The molbloc data header is corrupted.
See Also
3.2.2