molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company
Page 24
To determine if a molbox RFM is equipped with the microrange option, check the model
number on the product label on the bottom of the molbox RFM case. If the model number
includes a “-M”, the microrange option is installed.
The microrange option is only active during molbloc-L operation. Mirorange has no effect
on molbloc-S operation or measurement specifications.
3.1.6
REFERENCE PRESSURE TRANSDUCER (RPT)
OVERPRESSURE
3.1.6.1
UPSTREAM AND DOWNSTREAM ABSOLUTE RPTS
Every molbox RFM has two absolute RPTs, one measures molbloc upstream
pressure, the other usually measures molbloc downstream pressure and may
provide a second measurement of molbloc upstream pressure in molbloc-S
operation. In normal operation, they are not used at pressures greater than
600
kPa absolute (87 psia)
.
Exposing the molbox RFM RPTs to pressures greater than the maximum
operating pressure may damage them. molbox RFM has a system of warnings
and alarms to protect itself from overpressure (see Section 3.6.3.1).
3.1.6.2
DIFFERENTIAL RPT, MICRORANGE OPTION
The microrange option low differential RPT has no overpressure warnings or
alarms in normal operation. It is, to the extent possible, protected from
overpressure automatically and transparently as it is put into and out of service
using the molbox RFM on-board valving.
In
run calibration
operation (see Section 5.2.4.1), the microrange RPT has an
OVERPRESSURE function whose operation is similar to the absolute RPT
overpressure function (see Section 3.6.3). Overpressure occurs at about
13.7
kPa (2 psi)
differential.
When an overpressure condition occurs
during
run calibration
of the microrange
transducer, the display of microrange
pressure indicates:
Microrange RPT:
OVERP! kPa [ENTER]
Be sure the pressure conditions that caused the overpressure to occur have
been cleared. Then press
[ENTER]
to reactivate the microrange RPT.
When using molbox RFM equipped with the microrange option, avoid very rapid
pressurization of one molbloc port. Attempt to open flow isolation valves
slowly. The microrange option low differential RPT is protected from
overpressure automatically by internal valving. However, dumping pressure very
rapidly on one port of the molbloc while the differential RPT is active could
cause a very sudden surge in differential pressure that can overpressure it.