molbox™ RFM™ OPERATION AND MAINTENANCE MANUAL
© 1998-2007 DH Instruments, a Fluke Company
Page 116
OHMS
Purpose
Read the resistance of the active molbloc’s PRTs in ohms.
Syntax “OHMS”
Remarks
The most recent molbloc upstream and downstream PRT
resistance measurement is returned in ohms. An error will be
returned if there is a problem reading the molbloc PRTs.
Example Command:
Reply:
“OHMS”
“107.075 Ohms, 107.079 Ohms”
Errors
ERR#
42:
There is an error in the molbox RFM PRT
measurement.
See Also
3.2.5, “TEMP”, “STDRES”
PCAL(=)
Purpose
Read or set the user RPT calibration adjustments (
PA
,
PM
).
Syntax
“PCAL=upadder, upmult, dnadder, dnmult, ( mradder, mrmult)”
“PCAL”
Defaults
“PCAL=0, 1.00000, 0, 1.00000” (no microrange option)
“PCAL=0, 1.00000, 0, 1.00000, 0.0, 1.00000” (microrange option)
Arguments
upadder:
The upstream RPT calibration
adder
(
PA
) from –
10 000 to 10 000 Pa.
upmult:
The upstream RPT calibration
multiplier
(
PM
) from
0.9 to 1.1.
dnadder:
The downstream RPT calibration
adder
from –10
000 to 10 000 Pa.
dnmult:
The downstream RPT calibration
multiplier
from
0.9 to 1.1
mradder:
The optional microrange RPT calibration
adder
from -10 000 to 10 000 Pa.
mrmult:
The optional microrange RPT calibration
multiplier
from 0.9 to 1.1.
Remarks
The user defined pressure calibration for the upstream and
downstream RPTs can be access with this single command.
The
adder
is always in Pascal.
Example Command:
Reply:
“PCAL=38, 1.0021, 14, .9942” (no microrange
option)
“38 Paa, 1.00210, 14 Paa, 0.99420”
Command:
Reply:
“PCAL=38, 1.0021, 14, .9942, 6, 1.0001”
(microrange option)
“38 Paa, 1.00210, 14 Paa, 0.99420, 6.0 Paa,
1.0001”
Errors
ERR# 6:
Invalid argument.
See Also
5.2