4. REMOTE OPERATION
Page 125
© 1998-2007 DH Instruments, a Fluke Company
TARESET(=)
Purpose To
automatically
tare
the upstream and downstream RPT
(and microrange RPT if present) or set the
tare
to a user defined
value.
Syntax “TARESET=tare(,microrange
tare)
“TARESET”
Defaults “TARESET=0(,0)”
Arguments tare:
The
tare value
(-9999 to 9999 Pa).
microrange
tare:
The optional microrange transducer
tare
value (-
999 to 999 Pa).
Remarks
“TARESET” will automatically
tare
the molbox RFM only if the
tare
results are within the limits given (see the “TARE”
command). Use the “TARESETUP” command to prepare the unit
to be
tared
, and the “TARE” command to check if conditions are
valid. After you have completed taring the unit, use the
“TARESETUP” command or the “ABORT” command to return the
unit to normal operating conditions. If the microrange option is
installed, it is also
tared
. You can also manually set the
tare
value(s) using the “TARESET=tare” command. All pressure
values are always in Pascal [Pa]. While the
tare
cycle is
executing, a ‘b’ (“busy”) will appear in the third character position
of the “FR” and the “SR” command replies.
NOTE:
This command typically takes 5 to 7 seconds to complete
if performing a
tare
(no arguments given).
Example Command:
Reply:
Reply:
“TARESET”
“66 Pa”
“66 Pa, 7 Pa” (microrange option installed)
Errors ERR#
6:
ERR# 25:
An argument is invalid.
One of the
tare
values
exceed
tare
limits.
See Also
3.2.4.1, “TARE”, “TARESETUP”, “ABORT”
TEMP
Purpose
Read the temperature of the active molbloc in the current unit
of measure.
Syntax “TEMP”
Remarks
The upstream and downstream temperatures as well as average
molbloc temperature are returned in the current temperature
units. They are replied in the order of:
Average, upstream, downstream
An error will be returned if there is a problem reading the
molbloc PRTs.
Example Command:
Reply:
“TEMP”
“17.85 C, 17.84 C, 17.86 C ”
Errors
ERR#
37:
There is an error in the molbox RFM PRT
measurement.
See Also
3.2.5, “TUNIT”