4. REMOTE OPERATION
Page 115
© 1998-2007 DH Instruments, a Fluke Company
Example Command:
Reply:
Command:
Reply:
Reply:
Reply:
“MICRORNG=AUTO”
“AUTO”
“MICRORNG”
“OFF” (AUTO and MANUAL functions are
disabled)
“AUTO” (AUTO function is enabled)
“ON” (MANUAL function is enabled)
Errors ERR#
7:
ERR# 23:
The argument is invalid.
The microrange option is not installed in this
molbox RFM.
See Also
3.1.2.6, 3.2.7, and 3.4.3
MOLBLOC
Purpose
Read the active molbloc header data. This command does NOT
load molbloc information.
Syntax “MOLBLOC”
Remarks
The external molbloc header data includes the serial number,
range, calibration date, and PRT calibration data. This data is
stored in the molbloc, and is downloaded to the molbox RFM on
power up. The data for the active molbloc is returned in a single,
comma delimited string in the following order:
Serial Number
Range Designator
Range Flow Units
Gas
Calibration Date (yyyymmdd)
Total PRT Resistance (up downstream) at 0 °C [
Ω
]
PRT Slope [dimensionless]
Upstream PRT Resistance at 0 °C [
Ω
]
Downstream PRT Resistance at 0 °C [
Ω
]
Example Command:
Reply:
“MOLBLOC”
“100,1.00,sccm,N2,01/01/94,200.0050,0.7792,10
0.0010, 100.0040”
Errors ERR#
7:
ERR# 24:
ERR# 35:
ERR# 40:
ERR# 41:
The gas is invalid with the molbox RFM.
A molbloc is not connected to the molbox RFM.
The molbloc experienced a communications
timeout.
The molbloc experienced a read checksum
failure.
The molbloc data header is corrupted.
See Also
3.2.8, “DEVICE=EXT”