4. REMOTE OPERATION
Page 127
© 1998-2007 DH Instruments, a Fluke Company
See Also
3.3.3, “TEMP”
UDU(=)
Purpose
Read or set the user definable pressure unit.
Syntax “UDU”
“UDU=text, coef”
Arguments
text:
The text to use to identify the unit. It can be up to
4 characters wide. This will be the text used to
specify the unit if selecting it remotely, and that
appears on the molbox RFM.
coef:
The coefficient to convert Pascal [Pa] to the user
unit. The pressure (in Pa) will be multiplied by this
coefficient before being displayed.
Defaults “UDU=UNIT,
1”
Example Command:
Reply:
“UDU=Bar2, .00001”
“Bar2, 0.000010”
Errors ERR#
2:
ERR# 6:
The text argument is longer than 4 characters.
The coef argument is invalid.
See Also
3.3.2, “PUNIT”
USERCAL(=)
Purpose
Read or set the user flow measurement adjustment.
Syntax “USERCAL=
adder
, mult”
“USERCAL”
Defaults “USERCAL=0,
1”
Arguments
adder
: The
flow
adder
in the current flow unit and gas
type (0 to FS).
Mult:
The
flow
multiplier
(0.1 to 2).
Remarks
The user definable adjustment lets the user modify the measured
flow. The
adder
and
multiplier
are separated by a comma. The
adder
is in the current flow unit.
Example Command:
Reply:
“USERCAL=.58,.995”
“0.58 sccm, 0.995000”
Errors ERR#
6:
The
adder
or
multiplier
argument is invalid.
See Also
3.3.6, “PCAL(=)
UTEMP
Purpose
Read or set the temperature reference used for the volumetrically
based mass flow unit with user defined reference temperature.
Syntax “UTEMP=temp”
“UTEMP”
Defaults
“UTEMP=0” (
°
C)
Arguments temp:
The reference temperature in the current
temperature unit of measure.
Remarks
This temperature reference is used by the user defined units.
Example Command:
Reply:
“UTEMP=20”
“20.00 C”
See Also
3.2.3, 3.2.3.2, and 3.2.3.3, “FUNIT”, “TUNIT”