D-1
Appendix D
Specifications
Equipment Specifications
The following specifications apply to the Index 2MF and 2MFE.
%O2
Range ................................................................. 35 to 100 %
Resolution........................................................... 1 %
Accuracy ............................................................. 100 to 75 %: +1 %
±
accuracy of the pulse oximeter under test.
74 to 50 %: + 3 %
±
the accuracy of the pulse oximeter under test.
<50 %: unspecified
Repeatability ....................................................... +1 standard deviation
Rate
Range ................................................................. 30 to 250 BPM
Resolution........................................................... 1 BPM
Accuracy ............................................................. 1 % +1 BPM
Pulse Amplitude
Range ................................................................. 0 to 20 %
Resolution........................................................... 1 %
Battery Life
............................................................ At least 4 hours of continuous use
Battery Charger
Type .................................................................... Lead-Acid battery charger only
Input .................................................................... 100 – 250 V, 60 – 60 Hz, 0.3 A
Output ................................................................. 12 V dc, 0.5 A
Probe Test
............................................................. Index 2MFE version only
Continuity/Resistance Test
Test Matrix .......................................................... Measures all combinations of possible interconnections in an XX point
matrix.
Range ................................................................. 250
Ω
to 150 k
Ω
Accuracy .............................................................
5 % of reading
LED/Detector Voltage Test
Test Format ........................................................ Measures the voltage drop across Red LED, infrared LED, and the
photo detector when the internally generated test signal is applied.
Test Signal.......................................................... Constant current source at 1.0 mA
Open Circuit........................................................ 2.5 V max
Measurement/Display Range ............................. 0.0 to 4 V
Accuracy .............................................................
5 % of reading, 0.4 to 4 V
Summary of Contents for Biomedical Index 2MF
Page 8: ...Index 2MF Users Guide iv...
Page 10: ...Index 2MF Users Guide vi...
Page 12: ...Index 2MF Users Guide viii...
Page 18: ...Index 2MF Users Guide 1 6...
Page 28: ...Index 2MF Users Guide 4 4...
Page 34: ...Index 2MF Users Manual 5 6...
Page 72: ...Index 2MF Users Guide 9 6...
Page 76: ...Index 2MF Users Guide 10 4...
Page 86: ...Index 2MF Users Guide 12 4...
Page 88: ...Index 2MF Users Guide...
Page 94: ...Index 2MF Users Guide A 6...
Page 98: ...Index 2MF Users Guide C 2...
Page 112: ...Index 2MF Users Guide F 4...
Page 126: ...Index 2MF Users Guide H 6...