![Fluke Biomedical Index 2MF User Manual Download Page 59](http://html1.mh-extra.com/html/fluke/biomedical-index-2mf/biomedical-index-2mf_user-manual_2310044059.webp)
Using Test Programs
Creating a Custom Test Program (Autosequence)
8
8-9
esl094.eps
3.
Press
SEL
to change the
Include Actuals:
response from
YES
to
NO
as needed.
4.
Move to the next step in the process by pressing the softkey
ADV
.
Setting the 02 Level
You can set up to ten different oxygen levels and up to ten different pulse rates for the
simulation autosequences.
Note
In each screen, the softkey
ADV
moves ahead to the next settable option.
Use the
NEXT
softkey to simply advance through the O2 and rate level
settings for each test.
1.
After selecting a test, set the oxygen level for test number one:
esl095.eps
2.
Press
+
or
-
to adjust the simulated oxygen level up or down for the first O2 test
(1% O2 per button push). You can set up to ten O2 tests. Pressing
NEXT
at
Rate
Setting #10
brings you back to
O2 Setting #1
.
esl096.eps
3.
Press
NEXT
to set the simulated pulse rate.
Note
If you do not plan to test for O2, press
ADV
to bypass these settings.
Setting the Pulse Rate
You can set up to ten different pulse rates for the Simulator to use during a single
autosequence program.
1.
Again, use
+
and/or
-
to adjust the BPM up or down for the first pulse rate test.
Summary of Contents for Biomedical Index 2MF
Page 8: ...Index 2MF Users Guide iv...
Page 10: ...Index 2MF Users Guide vi...
Page 12: ...Index 2MF Users Guide viii...
Page 18: ...Index 2MF Users Guide 1 6...
Page 28: ...Index 2MF Users Guide 4 4...
Page 34: ...Index 2MF Users Manual 5 6...
Page 72: ...Index 2MF Users Guide 9 6...
Page 76: ...Index 2MF Users Guide 10 4...
Page 86: ...Index 2MF Users Guide 12 4...
Page 88: ...Index 2MF Users Guide...
Page 94: ...Index 2MF Users Guide A 6...
Page 98: ...Index 2MF Users Guide C 2...
Page 112: ...Index 2MF Users Guide F 4...
Page 126: ...Index 2MF Users Guide H 6...