![Fluke Biomedical Index 2MF User Manual Download Page 120](http://html1.mh-extra.com/html/fluke/biomedical-index-2mf/biomedical-index-2mf_user-manual_2310044120.webp)
Index 2MF
Users Guide
G-8
In some cases, it may be advantageous to
not
use the floating diode source. In these
cases, the Simulator provides a ground pin at pin 9, and both positive and negative
simulation voltage signals on pins 7 and 8. These may be converted to current signals by
appropriate scaling resistors as noted in Figure G-5.
Normally, oximeter shield pins should not be tied to the Simulator ground, but there may
be exceptions. Again, only a thorough study and understanding of the oximeter interface
circuits can answer these questions.
EXAMPLE: OHMEDA HARDWIRE CABLE
INDEX 2
PIN
OHMEDA PROBE
PIN
TO OXIMETER
IR CATHODE
R CATHODE
RED ANODE
15.8
ID RES
ID RES COMMON
DET. ANODE
SHLD/DET RET
T
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
X2
56K
1%
200K, 1%
100K, 1%
esl163.eps
Figure G-5. Example: Ohmeda Hardware Cable
Summary of Contents for Biomedical Index 2MF
Page 8: ...Index 2MF Users Guide iv...
Page 10: ...Index 2MF Users Guide vi...
Page 12: ...Index 2MF Users Guide viii...
Page 18: ...Index 2MF Users Guide 1 6...
Page 28: ...Index 2MF Users Guide 4 4...
Page 34: ...Index 2MF Users Manual 5 6...
Page 72: ...Index 2MF Users Guide 9 6...
Page 76: ...Index 2MF Users Guide 10 4...
Page 86: ...Index 2MF Users Guide 12 4...
Page 88: ...Index 2MF Users Guide...
Page 94: ...Index 2MF Users Guide A 6...
Page 98: ...Index 2MF Users Guide C 2...
Page 112: ...Index 2MF Users Guide F 4...
Page 126: ...Index 2MF Users Guide H 6...