5-1
Chapter 5
Configuring the Simulator
Configuring for Specific Oximeter Make
Light technology and R-Curve data must be known in order to test a pulse oximeter,
optically via finger or electrically. You can configure the Simulator to match the make of
pulse oximeter(s) you will be testing. The Simulator stores the definitions for different
makes of pulse oximeters. You can change variables for each of the pulse oximeter types
to match the devices you will be testing.
Accessing the Make Menu
Change the make of the pulse oximeter stored in the Simulator's memory using the Make
menu.
1.
To get to the Make menu, press
more
from Main Menu 1.
fgg014.eps
2.
You will then see the screen for the Main Menu 2. Press the softkey
MAKE
.
fgg169.eps