Testing the Pulse Oximeter Limits
Simulating the Pulse Rate
7
7-3
esl054.eps
The results will be sent out the RS232 and the Centronics ports to an attached printer.
Note
You can connect the Simulator to any printer with a serial or parallel
Centronics-type port or to any IBM-compatible PC, including laptops and
notebooks. For information on how to connect the Simulator to other
systems or peripherals, see “Appendix A: Printing & Data Transfer”.
If a printer or computer is attached, your results should look like this example:
O2 Alarm Response Time:
-------------------------
13.0
sec
5.
When you are finished testing the pulse oximeter for SpO2 readings, press
esc
to
return to the Limits menu.
esl055.eps
Simulating the Pulse Rate
The Simulator simulates a variety of patient conditions, while you observe the effect of a
changing pulse rate with a static SpO2 on the pulse oximeter, as well as the effects of
using a changing pulse rate with a changing SpO2 percentage.
You can vary the pulse rate (beats per minute) of the simulated patient index finger or
simulate the pulse rate electrically.
1.
To access the pulse rate alarm menu, press the softkey
LMTS
from Main Menu 1,
and then press
BPM
to access the Rate Alarm menu.
esl056.eps
You will see this pulse Rate Alarm menu displayed on the LCD:
Summary of Contents for Biomedical Index 2MF
Page 8: ...Index 2MF Users Guide iv...
Page 10: ...Index 2MF Users Guide vi...
Page 12: ...Index 2MF Users Guide viii...
Page 18: ...Index 2MF Users Guide 1 6...
Page 28: ...Index 2MF Users Guide 4 4...
Page 34: ...Index 2MF Users Manual 5 6...
Page 72: ...Index 2MF Users Guide 9 6...
Page 76: ...Index 2MF Users Guide 10 4...
Page 86: ...Index 2MF Users Guide 12 4...
Page 88: ...Index 2MF Users Guide...
Page 94: ...Index 2MF Users Guide A 6...
Page 98: ...Index 2MF Users Guide C 2...
Page 112: ...Index 2MF Users Guide F 4...
Page 126: ...Index 2MF Users Guide H 6...