Index 2MF
Users Manual
7-2
esl049.eps
The display changes to present these options:
esl051.eps
When setting new oxygen limits, the Simulator displays the last setting used (or 96%
if not previously changed) and resets the time to
00.0
seconds.
2.
Press the softkeys
+
or
-
to start the timer, and the screen displays the “Timing!”
message.
esl052.eps
•
Pressing the softkey
+
(plus) will
increase
by the step amount set previously
using the Simulations menu without affecting the timing function.
•
Pressing the softkey
–
(minus) will
decrease
by the step amount without
affecting the timing function.
Note
Continue pressing
+
or
-
until the O2 percent you want to test for is
displayed. Each press pf the softkey
+
or
–
resets the timer internally and
restarts the timing.
3.
If the O2 setting is outside the oximeter's alarm limit, the alarm will sound on the
pulse oximeter being tested after a period of time. Press
STOP
immediately. This
stops the timer, and displays the “elapsed time to alarm” for the specified simulation
level.
esl053.eps
4.
To print the results of the alarm, press
PRT
.
Summary of Contents for Biomedical Index 2MF
Page 8: ...Index 2MF Users Guide iv...
Page 10: ...Index 2MF Users Guide vi...
Page 12: ...Index 2MF Users Guide viii...
Page 18: ...Index 2MF Users Guide 1 6...
Page 28: ...Index 2MF Users Guide 4 4...
Page 34: ...Index 2MF Users Manual 5 6...
Page 72: ...Index 2MF Users Guide 9 6...
Page 76: ...Index 2MF Users Guide 10 4...
Page 86: ...Index 2MF Users Guide 12 4...
Page 88: ...Index 2MF Users Guide...
Page 94: ...Index 2MF Users Guide A 6...
Page 98: ...Index 2MF Users Guide C 2...
Page 112: ...Index 2MF Users Guide F 4...
Page 126: ...Index 2MF Users Guide H 6...