6-1
Chapter 6
Setting and Changing Simulation Factors
About Simulations and Presets
One of the main functions of the Simulator is to test the performance of a pulse oximeter
for any possible patient condition. The Simulator allows you to set any combination of
blood-oxygen levels, heart rates, and pulse amplitude, as well as simulate a virtual range
of medical physical conditions and set an ambient light artifact.
Setting this group of default factors allows you to test an oximeter and determine its
weaknesses, if any. The Simulator will display and optionally print areas in your pulse
oximeter performance that may be unacceptable, and document correct performance.
The simulations set here are used as the defaults when testing the pulse oximeter.
Setting Simulations
To access the Simulations Menu, press the softkey
SIM
from the Main Menu.
fgg010.eps
The Simulations Menu
The Simulations menu appears in the LCD with these choices:
•
MAN
to manually set the O2 and BPM
•
PRESET
to set a patient’s health condition
•
AMB
to set an ambient light artifact
•
STEP
to change step size for saturation and BPM
•
esc
to return to previous menu (Main Menu 1)
Summary of Contents for Biomedical Index 2MF
Page 8: ...Index 2MF Users Guide iv...
Page 10: ...Index 2MF Users Guide vi...
Page 12: ...Index 2MF Users Guide viii...
Page 18: ...Index 2MF Users Guide 1 6...
Page 28: ...Index 2MF Users Guide 4 4...
Page 34: ...Index 2MF Users Manual 5 6...
Page 72: ...Index 2MF Users Guide 9 6...
Page 76: ...Index 2MF Users Guide 10 4...
Page 86: ...Index 2MF Users Guide 12 4...
Page 88: ...Index 2MF Users Guide...
Page 94: ...Index 2MF Users Guide A 6...
Page 98: ...Index 2MF Users Guide C 2...
Page 112: ...Index 2MF Users Guide F 4...
Page 126: ...Index 2MF Users Guide H 6...