![Fluke Biomedical Index 2MF User Manual Download Page 68](http://html1.mh-extra.com/html/fluke/biomedical-index-2mf/biomedical-index-2mf_user-manual_2310044068.webp)
Index 2MF
Users Guide
9-2
Note
The MFE version ships with Nellcor and Ohmeda probe and oximeter
adapters. Please contact Fluke Biomedical regarding the availability of
additional adapter cables.
W
Caution
Attempting to create adapter cables without all details of the
oximeter probe-operating scheme could result in damage to the
oximeter, the Simulator, or both.
2.
Press the
PRBE
softkey at Main Menu 1:
Fgg012.eps
The
PRBE
softkey initiates a testing sequence that includes:
•
Verifying that a Probe Test Board is installed in the Simulator
•
Checking whether or not a probe adapter cable is connected to the probe port
•
Performing all measurements for LED, photo detector, and resistance tests
esl125.eps
Probe Test Selection
The user can then choose the specific probe test.
•
LED and photodiode electrical test
•
Pin-to-pin resistances
•
Photodiode detector optical test
Note
During the photodiode test, the finger probe being tested should not be
attached to the Index finger.
LED Testing
The Probe test should end with the following menu displayed. Press the
LEDS
softkey:
Summary of Contents for Biomedical Index 2MF
Page 8: ...Index 2MF Users Guide iv...
Page 10: ...Index 2MF Users Guide vi...
Page 12: ...Index 2MF Users Guide viii...
Page 18: ...Index 2MF Users Guide 1 6...
Page 28: ...Index 2MF Users Guide 4 4...
Page 34: ...Index 2MF Users Manual 5 6...
Page 72: ...Index 2MF Users Guide 9 6...
Page 76: ...Index 2MF Users Guide 10 4...
Page 86: ...Index 2MF Users Guide 12 4...
Page 88: ...Index 2MF Users Guide...
Page 94: ...Index 2MF Users Guide A 6...
Page 98: ...Index 2MF Users Guide C 2...
Page 112: ...Index 2MF Users Guide F 4...
Page 126: ...Index 2MF Users Guide H 6...