Index 2MF
Users Manual
7-6
esl063.eps
3.
To raise the Pulse Amp percentage (AC factors) in the simulated blood condition,
press the softkey
+
:
esl064.eps
The amplitude will
increase
in increments of 1% with every softkey press.
4.
To lower the percentage of AC factors in the simulated blood condition, press the
–
softkey:
esl065.eps
The amplitude will
decrease
by 1% with every softkey press down to 1 %. From 1 %
to 0.1 % the value will decrease by 0.1 %. Below 0.1 %, the value will decrease by
0.025 % per button press. Inputs below 6 % are rounded internally to the nearest
0.024 %, and above 6 % are rounded to the nearest 0.072 %. The displayed value is
the rounded value nearest the value displayed per the preceding increment/decrement
rules. If the pulse amplitude rate exceeds the oximeter alarm rate, the alarm will
sound on the pulse oximeter being tested after a period of time.
5.
To print the results of the alarm, press the
PRT
softkey.
esl067.eps
The results will be sent to the RS232 or Centronics port.
Summary of Contents for Biomedical Index 2MF
Page 8: ...Index 2MF Users Guide iv...
Page 10: ...Index 2MF Users Guide vi...
Page 12: ...Index 2MF Users Guide viii...
Page 18: ...Index 2MF Users Guide 1 6...
Page 28: ...Index 2MF Users Guide 4 4...
Page 34: ...Index 2MF Users Manual 5 6...
Page 72: ...Index 2MF Users Guide 9 6...
Page 76: ...Index 2MF Users Guide 10 4...
Page 86: ...Index 2MF Users Guide 12 4...
Page 88: ...Index 2MF Users Guide...
Page 94: ...Index 2MF Users Guide A 6...
Page 98: ...Index 2MF Users Guide C 2...
Page 112: ...Index 2MF Users Guide F 4...
Page 126: ...Index 2MF Users Guide H 6...