Testing the Pulse Oximeter Limits
Simulating Pulse Amplitude
7
7-5
esl060.eps
The results will be sent to the RS232 and Centronics port.
Note
Connect the Simulator to any printer with a serial port or parallel
Centronics port, or to any IBM-compatible PC, including laptops and
notebooks. For information on how to connect to other systems or
peripherals, see “Appendix A: Printing & Data Transfer”.
If a printer or computer is attached, you will see results that look something like this
example:
Rate Alarm Response Time
-------------------------
10.0 sec
Press
esc
to return to the Limits menu.
esl061.eps
Simulating Pulse Amplitude
The peak-to-peak amplitude of the simulated blood pressure wave can be increased or
decreased. Decreasing amplitude corresponds to a weakening pulse. You can decrease
amplitude to find where the oximeter fails to find the pulse.
You can quickly set a pulse amplitude limit.
1.
To access the pulse amplitude alarm menu, press
LMTS
from Main Menu 1, and then
AMP
from the Limits menu:
esl062.eps
2.
The Pulse Amplitude menu displays in the LCD screen:
Summary of Contents for Biomedical Index 2MF
Page 8: ...Index 2MF Users Guide iv...
Page 10: ...Index 2MF Users Guide vi...
Page 12: ...Index 2MF Users Guide viii...
Page 18: ...Index 2MF Users Guide 1 6...
Page 28: ...Index 2MF Users Guide 4 4...
Page 34: ...Index 2MF Users Manual 5 6...
Page 72: ...Index 2MF Users Guide 9 6...
Page 76: ...Index 2MF Users Guide 10 4...
Page 86: ...Index 2MF Users Guide 12 4...
Page 88: ...Index 2MF Users Guide...
Page 94: ...Index 2MF Users Guide A 6...
Page 98: ...Index 2MF Users Guide C 2...
Page 112: ...Index 2MF Users Guide F 4...
Page 126: ...Index 2MF Users Guide H 6...