SELECT A DEVICE TO CALIBRATE
>SPRT 1
TCS
2
PRTS 3
GPIB 4
PRNT 4
AOUT 4
$
After the device is selected, another window appears showing the parameters
and functions available from the device. These depend on the device. The de-
vice calibration window might appear as follows:
CALIBRATE DEVICE: SPRT
0 ADJ: 0.0
100 ADJ: 0.0
400 ADJ: 0.0
CAL DATE: 05-21-96
SER NUM: 123456
If there are no calibration commands for a particular device the window shows
“NO CALIBRATION OPTIONS FOR THIS DEVICE.”
You can scroll through the calibration parameters using the
UD
buttons. The
values of parameters can be changed using the numeric keys or the
LR
but-
tons as is appropriate for the parameter type. Press
ENTER
to set the
parameter.
With function commands you can use the
LR
buttons to change “NO” to
“YES” and then press
ENTER
to execute the function. If
ENTER
is pressed
with “NO” nothing will happen except the cursor will move down to the next
line. Press
EXIT
to exit.
5.4.4
Module Information
The
MODULE INFO
soft-key function displays information about a module.
This includes the module name, model number, firmware version number, and
the names of devices it contains with the class and number of channels for each
device (remember a module may contain more than one independent device).
First, you must select the module. A window appears showing a list of mod-
ules. Each module is followed by the module number. The module is selected
using the
UD
buttons and pressing
ENTER
.
1560 Thermometer Readout
User’s Guide
56