and the probe serial number. The channel is selected using the
UD
buttons
and pressing
ENTER
.
SELECT A CHANNEL TO EDIT
>SPRT 01, PROBE 566-011
SPRT 02, PROBE 566-012
TCS
03, PROBE 1341
TCS
04, PROBE 1342
TCS
05, PROBE 1343
TCS
06, PROBE 1344
$
After the channel is selected, a new window appears allowing you to edit the
probe parameters for the selected channel. The probe parameters include the
probe serial number, conversion type, and characterization coefficients and pa-
rameters. The available coefficients may change depending on the conversion
type that is selected.
EDIT SPRT 01
PROBE SER#: 566-011
CONVERSION: ITS-90
LO RANGE: 4
HI RANGE: 7
RTPW: 25.546738
A[4]: -1.5763669E-4
$
The PROBE SER# parameter is the serial number for the probe. The serial
number consists of a string of up to eight characters using any numeric digits,
letters, minus signs, and decimal points. Use the appropriate numeric buttons to
enter digits. Letters are entered by pressing the
EXP
button. Press this button
repeatedly until the desired letter appears.
The CONVERSION parameter specifies the conversion type. The entire list of
possible conversion types is given in Table 3. The conversion types available
for a given channel depend on the type of input as indicated in the table. The
conversions and related coefficients and parameters are explained in the follow-
ing sub-sections.
1560 Thermometer Readout
User’s Guide
34
Table 3
Conversion Types