ENABLE PRINTING TO DEVICES
>SERI 0: ON
LR
PRNT 4: ON
Each time a new measurement is produced and displayed on the screen in the
primary output window it is also sent to all enabled printer devices. Each mea-
surement is printed with the channel number, measurement value, unit prefix,
unit, time (in 24-hour format), and date.
5.3.4
Print Memory
The
PRINT MEMORY
soft-key prints measurements stored in memory to any
printer output device. Up to 1000 of the most recent measurements can be
printed. (The storage capacity of the memory may be reduced if many input
channels are added.) You are requested to select a printer device. You can scroll
through the list using the
UD
buttons. The indicated printer device is selected
by pressing
ENTER
.
SELECT PRINTER DEVICE
>SERI 0
SERC 4
Next, you must enter the number of measurements to print. Use the numeric
buttons to enter a value from 1 to 1000 and press
ENTER
.
1560 Thermometer Readout
User’s Guide