2 .1 .7 Condensate deposits
Oil or water condensates on the sensors may falsify the measuring results. Such deposits must be
expected, for example, when high flow velocities of compressed air are concerned and there is no
drying provided. Normal condensation on an unheated sensor is normally not realised, it will dry up
after a few minutes. Deposits on heated sensors only occur at high air humidities and will cause
significant measuring errors. In most cases, such deposits will dry up within a few minutes. Deposits
of oil, however, will not dry up and should be removed at regular intervals.