ERRORS
59
FC
01- CA
| Flow Meter
8 .2 Potential errors
Independent of the priority group, all errors found are indicated with their relevant number.
In order to facilitate operation, the last error is stored in a non-volatile memory. The stored error
may be retrieved and deleted in the main menu.
If a combination of errors occurs simultaneously, they are indicated or stored in the error memory
observing the following sequence.
Priority group I
Error
Cause
Rectification
No. 1
No system parameter available
Return to supplier.
No. 2
Incorrect test sum of parameter
Return to supplier.
memory
No. 3
Incorrect test sum of program
Return to supplier.
memory
No. 4
Incorrect test sum of data memory
Return to supplier.
No. 5
Internal controller error
Return to supplier.
Priority group II
Error
Cause
Rectification
No. 10
Sensor not connected; or cable
Check cable or replace sensor.
between FC01-CA and sensor
defective; or defective sensor
Sensor selected (configuration)
Correct sensor selection in
differs from sensor connected
configuration menu
No. 21
Medium temperature too high