9 .3 .1 Voltage output V1 - 5 V FS
Signal voltage range:
U
S
= 0 V (1 V) to 5 V ±2 % FS
Max. signal ripple:
dU
S
= 5 % FS
Min. admissible load resistance:
R
l
= 1 k
Ω
Max. admissible load capacity:
C
l
= 1 nF
Max. admissible load inductance:
L
l
= 100 nH
Short circuit proof:
yes (XAO - between all terminals)
9 .3 .2 Voltage output V2 - 10 V FS
Signal voltage range:
U
S
= 0 V (2 V) to 10 V ±2 % FS
Max. signal ripple:
dU
S
= 5% FS
Min. admissible load resistance:
R
l
= 2 k
Ω
Max. admissible load capacity:
C
l
= 1 nF
Max. admissible load inductance:
L
l
= 100 nH
Short circuit proof:
yes (XAO - between all terminals)
9 .3 .3 Current output C1 - 20 mA FS
Signal current range:
I
S
= 0 mA (4 mA) to 20 mA ±2 % FS
Max. signal ripple:
dI
S
= 5% FS
Min. admissible load resistance:
R
l
= 0
Ω
Max. admissible load resistance:
R
l
= 250
Ω