BE6000 Ultrasonic Flow Meter O&M Manual
Flomotion Systems, Inc. ~ FlomotionSystems.com
34
•
Flow Batch Controller (M81)
The internal batch controller in the system is able to control the input signals through keypad or analog input.
Output signals can be transmitted through OCT or relay. The flow batch value can be modified in this
window. The screen will enter the batch control display as soon as the modification is completed.
•
Date Totalizer (M82)
In this window, it is possible to review the historical flow data totalizer for any day for the last 64 days, any
month for last 64 months and any year for last 5 years. Press ENT, use UP or DOWN key to review totalizer
in days, months and years. Use UP or DOWN key to review the flow total for a specific day, month or year.
For instance, to display the flow total for May 17, 2005, the display “-----------“ at the upper right corner of the
screen indicates that it was working properly the whole day. On the contrary, if “G” is displayed, it indicates
that the instrument gain was adjusted at least once. Probably it was offline once on that day. If “H” is
displayed, it indicates that poor signal was detected at least once. Also, it indicates that the operation was
interrupted or problems occurred in the installation.
For details, please refer to Chapter 5 – Error Diagnoses
0. Day
1. Month
2. Year
•
Automatic Flow Correction (M83)
With the function of automatic flow correction, the flow lost in an offline session can be estimated and
automatically adjusted. The estimate is based on the average value, which is obtained from flow rate before
going offline and flow measured after going online the next time, multiplied times the time period that the
meter was offline. Select “No” to cancel this function.
•
Energy Unit Selection (M84)
Select GJ or K.C as energy unit. The default is GJ.
•
Energy Temperature Source Selection (M85)
Energy Unit Select [84
0. Giga Joule(GJ)
Automatic Correction
ON
Date Totalizer [82
0. Day
Flow Batch Controller
10000 x 1m