BE6000 Ultrasonic Flow Meter O&M Manual
Flomotion Systems, Inc. ~ FlomotionSystems.com
31
•
AI4 Range(M66)
To input scale range of temperature, pressure or level for the analog input of 4-20mA. For example, the
value 10 in the above box represents 4mA and 100 20mA.
•
Frequency Output Signal Frequency Range(M67)
To input low(12-9999Hz) and high(12-9999Hz) frequency for the output. Make sure to set OCT to output
frequency signal in order to output frequency signal.
•
Flow Value of Low Frequency Output(M68)
To input flow rate value for the low frequency point.
•
Flow Value of High Frequency Output(M69)
To input flow rate value for the high frequency point.
•
LCD Back Light Controller (M70)
LCD can be controlled using this window. “Always On” means the back light will be always on. “Always Off”
means always off. “Light for nn” means it will turn on when a key is pressed and will stay on for nn seconds
after the keys have been pressed. This option is provided for power saving.
•
LCD Contrast Controller (M71)
To control the LCD contrast. Press ENT and adjust the number using up or down keys, and press ENT again
to confirm.
•
Working Timer (M72)
High FO Flow Rate [69
0 m
3
/h
LCD Backlit Option [70
1. Always On
LCD Contrast [71
9
Low FO Flow Rate [68
0 m
3
/h
AI Value Range [66
10-100
FO Frequency Range [67
1-1001