BE6000 Ultrasonic Flow Meter O&M Manual
Flomotion Systems, Inc. ~ FlomotionSystems.com
13
3. Rapid Setup Procedure
(1) Press [MENU] [1] [1] to enter M11 window, input pipe outer diameter, press [ENT].
(2) Press [
▼
] to enter M12 window, input wall thickness and press [ENT].
(3) Press [
▼
] [
▼
] to enter M14 window and [ENT] [
▼
] [
▲
] to select pipe material, and press [ENT].
(4) Press [
▼
] to enter M16 window and [ENT] [
▼
] [
▲
] to select liner material, and press [ENT].
(5) Press [
▼
] to enter M20 window and [ENT] [
▼
] [
▲
] to select fluid type, and press [ENT].
(6) Press [
▼
] to enter M23 window and [ENT] [
▼
] [
▲
] to select the transducer type, and press [ENT].
(7) Press [
▼
] to enter M24 window and [ENT] [
▼
] [
▲
] to select the transducer mounting method, and
press [ENT].
(8) Press [
▼
] to enter M25 window, which will display the correct transducer mounting distance(the
distance between the front edges of the transducers).
(9) Press [MENU] [0] [1] [ENT], the flow rate and the velocity of the flow will be displayed upon
completion of the gain adjusting process.
4. How to Go into a Specific Window
The user can enter into specific windows using the [MENU] key followed by a 2-digit code. So, for example,
in order to go to window M13, which is the window for inputting the inner diameter of the pipe, press [MENU]
[1] [3] [ENT].
5. Transducer Mounting Distance
The distance is measured between the front edges of both transducers. After all the parameters have been
set, the distance is shown in M25.