![Flomotion Systems be6000 series Installation And Operation Manual Download Page 22](http://html1.mh-extra.com/html/flomotion-systems/be6000-series/be6000-series_installation-and-operation-manual_2306731022.webp)
BE6000 Ultrasonic Flow Meter O&M Manual
Flomotion Systems, Inc. ~ FlomotionSystems.com
22
You can select one of the following fluids from the list provided;
0.
Water
8.
Other
1. Sea Water
9. Diesel Oil
2. Kerosene
10. Caster Oil
3. Gasoline
11. Peanut Oil
4. Fuel Oil
12. Gasoline #90
5. Crude Oil
13. Gasoline #93
6. Propane (-45°C)
14. Alcohol
7. Butane (0°C) 15.
Water
(125°C)
If ‘8. Other’ is selected; input the corresponding sound velocity of the fluid in M21.
•
Fluid Sound Velocity(M21)
Enter the sound velocity of the fluid you chose in M20. This is activated only when ‘8. Other’ is selected in
M20.
•
Fluid Viscosity(M22)
Enter the viscosity of the fluid you chose in M20. This is activated only when ‘8. Other’ is selected in M20.
•
Transducer Type(M23)
Select one of the transducer types from the list provided;
0. Standard – M
4. Standard - B
1. Plug-in Type c
5. Plug-in Type B45
2. Standard – S
6. Standard – L
3. User’s Own Type
If you select ‘3. User’s Own Type, you should enter a group of transducer parameters such as sound wedge
angle, sound wedge velocity, ultrasonic delay and distance between the edge of
the transducer and sound.
•
Transducer Mounting Method(M24)
Fluid Sound Velocity
1600.0m/s
Fluid Viscosity
1.0054 cSt
Transducer Mounting
0. V
Transducer Type
0. Standard – M
Fluid Type
0. Water