6.8 Trigger IN
General
Trigger IN is used to tag images on the camera side so they are recorded on the software.
The status of the trigger IN signal at start of integration is added on the frame header sent to
the recording software.
∆
T is a jitter of one frame period maximum. The Trigger IN signal must be at least one frame
period long. All frames are sent to the PC. ResearchIR Max will start or stop acquisition
based upon Trigger IN signal. This is configured in the start and stop conditions of the
ResearchIR Max recording tab of the left panel.
When to use this configuration?
-
To capture a fugitive event. The camera will acquire images, but the software only records
the frame of interest.
-
When a precise start of recording time is requested.
-
When only few frames are needed to be recorded
When NOT to use this configuration?
-
When it is needed to trigger each acquisition. In that situation, it is preferable to use the Sync
IN input.
Characteristics
Name
Value
Amplitude (V)
Rising Edge TTL 0/+5 V
High state minimum voltage
>3.5 V
Low state maximum voltage
<0.5 V
Minimum Pulse width (ns)
300
Impedance
User selectable. 50 Ohms / 10 MOhms
Protection
Voltage peaks (500
V / < 1
ns)
Overvoltage (15 V)
Reversed polarity
Connector type
Coaxial BNC Jack
LED Description
LED Status
Description
OFF
No signal detected. Check connection and
signal levels
GREEN
Signal is detected and signal voltage is
correct, but signal continuous
ORANGE
Signal is detected.
Signal voltage is incorrect
Blinking Green
Signal is detected.
LED is blinking at signal frequency.
Signal voltage is correct
Blinking Orange
Signal is detected.
LED is blinking at signal frequency.
Signal voltage is incorrect