7. Aim the camera toward a point of interest. The object should be within the frame of
the measurement box.
8. To reset the sampled average, push and hold the programmable button
.
9. To sample, push the programmable button
.
10. Aim the camera toward more points of interest. Sample 10 times to build up a sample
base by pushing the programmable button
.
The alarm is now set up and ready to use. Occasionally record a few samples if the
alarm is used for a long time or if the conditions change.
Note
• Every time you push the programmable button
, a sample is saved. Make sure you
aim the camera toward an object within the interesting temperature range when you
push (and hold) the button.
• The algorithm has a memory of the last 10 samples. It discriminates between the high-
est and lowest values, and calculates an average of the remaining values.
• Do not modify the measurement setup or activate another alarm because this will de-
activate the screening alarm.