2. Select
(
Color
) and push the joystick. This displays a submenu.
3. Use the joystick to select a different palette.
4. Push the joystick to confirm and exit the menu mode.
16.6
Changing the measurement parameters
For accurate measurements, it is important to set the measurement parameters:
•
Emissivity
.
•
Reflected temperature
.
•
Object distance
.
•
Atmospheric temperature
.
•
Relative humidity
.
•
External IR window compensation
.
Emissivity
is the most important measurement parameter to set correctly. If the
Emissivity
is set to a low value, the
Reflected temperature
also becomes important. The parameters
Object distance
,
Atmospheric temperature
, and
Relative humidity
are relevant for longer
distances. The
External IR window compensation
must be activated if a protective win-
dow or external lens is used.
You can set the measurement parameters globally. You can also change the
Emissivity
,
Reflected temperature
, and
Object distance
parameters locally for a measurement tool.
For more information, see section 18.5
Changing the measurement parameters
16.7
Performing a non-uniformity correction
(NUC)
16.7.1
General
When the thermal camera displays
Calibrating...
it is performing what in thermography is
called a ”non-uniformity correction” (NUC). An NUC is
an image correction carried out by
the camera software to compensate for different sensitivities of detector elements and
other optical and geometrical disturbances
3
. For more information, see section 34
, page 124.
An NUC is performed automatically, for example at start-up, when changing a measure-
ment range, or when the environment temperature changes.
You can also perform an NUC manually. This is useful when you have to perform a critical
measurement with as little image disturbance as possible. You may, for example, want to
perform a manual calibration just before you start recording a video sequence.
16.7.2
Performing an NUC manually
16.7.2.1
Procedure
Follow this procedure:
1. To perform a manual NUC, push and hold down the image archive button
for
more than 2 seconds.
Note
You can also assign the function
Calibrate
to one of the programmable buttons.
For more information, see section 13.11
Assigning functions to the programmable but-
, page 36.
#T559954; r. AT/65353/65353; en-US
66
3. Definition from the European standard EN 16714-3:2016, Non-destructive Testing—Thermographic Testing—
Part 3: Terms and Definitions.
Summary of Contents for 72501-0101
Page 2: ......
Page 3: ......
Page 5: ...User s manual FLIR T10xx series T559954 r AT 65353 65353 en US v ...
Page 6: ......
Page 14: ......
Page 43: ...Screen elements 11 T559954 r AT 65353 65353 en US 29 ...
Page 125: ...Mechanical drawings 30 See next page T559954 r AT 65353 65353 en US 111 ...
Page 134: ...Digital I O connection diagram 32 See next page T559954 r AT 65353 65353 en US 120 ...
Page 135: ......
Page 136: ...CE Declaration of conformity 33 See next page T559954 r AT 65353 65353 en US 122 ...
Page 137: ......
Page 144: ... T559954 r AT 65353 65353 en US 130 ...
Page 145: ......