42
CHEMPURE MANUAL
SECTION 13
APPENDIX
Method: Hydrazine -02A (0-200 ppb)
HYDRAZINE IN BOILER WATER
RANGE: 0-200 ppb (ug/l)
GENERAL DESCRIPTION
The automated procedure for the determination of
hydrazine is based on the reaction with pdimethy-
laminobenzaldehyde in an acid medium to form p-
dimethylaminobenzaldehyde hydrazone. The yellow
color is measured at 460 nm.
* Watt, G.W. and Chrisp. J.D. Analytical Chemistry, 24,
2006
REAGENTS
REAGENT #1, ACID REAGENT (Red), #180-1018-31
REAGENT #2, DYE (DMAB), (Orange), #180-1018-32
WORKING STANDARDS
200 ppb (ug/l), #180-1018-35
OPERATING NOTES
1. Where Particulate matter is present, the solution
must be filtered prior to the determination. The
sample stream must be clarified before being intro-
duced to the system with an on-line filter
such as Rosemount Analytical’s IF-100 (part #180-
1400-01).
2. Flowcell: 12mm
#180-M243-12
Filter: 460 nm
#214-B224-13
Harness Assembly:
180-S001-17
Reaction vessel assembly:
#180-S002-30
Analog Board
#180-B025-02
(low wavelength):
Method: Phosphate-02A (0-3 mg/l, 0-5 mg/l, 0-10
mg/l)
PHOSPHATE IN BOILER WATER
(RANGE: 0-3 mg/l, 0-5 mg/l, 0-10 mg/l P)
GENERAL DESCRIPTION
The automated procedure for the determination of sol-
uble phosphate depends on well-known chemistry
whereby ammonium molybdate reacts in an acid
medium to form molybdophosphoric acid. This is sub-
sequently reduced to the molybdenum blue complex
by reaction with ascorbic acid.
REAGENTS
REAGENT #1 ASCORBIC ACID, (Red), #180-1030-34
plus #180-1030-36
REAGENT #2 AMMONIUM MOLYBDATE, (Orange),
#180-1047-35
STANDARDS
Working standard, 3.0 ppm, #180-1047-43
Working standard, 10.0 ppm. #180-1047-38
Working standard, 5.0 ppm, 3180-1047-39
OPERATING NOTES
1. The ammonium molybdate must be made up in
transparent flasks. The reagent should be clear. If
it has any blue color, it should be rejected. Also, it
should be free of precipitates
after a peri-
od of standing.
2. The sample stream must be clarified before being
introduced to the system with an on-line filter such
as Rosemount Analytical’s IF-100 (part #180-1400-
01).
3. Flowcell: 6mm
#180-M243-06
Filter: 660nm
#214-B224-53
Harness Assembly:
#180-S001-17
Reaction Vessel Assembly:
#180-S002-30