6
CHEMPURE MANUAL
SECTION 3
INSTRUMENT DESCRIPTION
3.3.2 CPU BOARD
Part
Function
20 mA Adjust
For adjustment of fullscale
Set Screw
recorder output signal when
recorder has no span adjust-
ment.
4 mA Adjust
For adjustment of zero
Set Screw
recorder output signal when
recorder has no span adjust-
ment.
5 V Adjust
For adjustment of fullscale
Set Screw
recorder output signal when
recorder has no span adjust-
ment.
0 V Adjust
For adjustment of zero
Set Screw
recorder output signal when
recorder has no span adjust-
ment.
Computer Memory
Clears microprocessor
Reset Switch
memory.
3.3.3 FAULT/ANALOG BOARD
Part
Function
Leak Detector
Adjusts sensitivity of leak
Adjust Set Screw
detector.
FIGURE 3-4 CPU BOARD
FIGURE 3-5 FAULT/ANALOG BOARD