SECTION 4
INSTALLATION
SECTION 4
INSTALLATION
After the analyzer has been unpacked and all parts
accounted for, install as described in the following
chapter.
4.1 MAIN CABINET MOUNTING
4.1.1 GENERAL RECOMMENDATIONS:
•
Mount the cabinet on a wall or other suitable verti-
cal surface, at least 66" (168cm ) wide and 43 1/2"
(111cm ) off the floor.
• Allow two feet of clearance at the left side of the
analyzer for access to the input/output panel, and
allow for swing of main and reagent cabinet doors.
• Allow 2 feet at the right side for access to the filter
and overflow tube assembly.
• Mount cabinet level or with a slight backward tilt for
proper leak detector operation.
• Mounting surface must be able to support 200
pounds.
•
If the analyzer is to be installed out of doors, it must
be suitably sheltered to maintain the following envi-
ronmental conditions: Temperature: 50°F to 90°F,
regulated ±10°F Humidity:
10 to 80%.
To mount the main cabinet, proceed as follows:
1. Drill 6 holes in wall for mounting brackets, as indi-
cated in
Figure 4-1
.
2. Securely attach “L” brackets to wall using suitable
hardware. (Brackets must be anchored properly to
support 200 pounds.)
3. Attach main cabinet to “L” brackets using four
3/8" x 3/4" bolts, 3/8" split ring washers, and 3/8"
hex nuts.
4.2 REAGENT CABINET MOUNTING
To mount the reagent cabinet, proceed as follows:
1. Loosen the four bolts in bottom of
main cabinet,
leaving approximately 1/4" between washer and
cabinet.
2. Lift reagent cabinet and attach to main cabinet by
placing heads of bolts into the keyhole slots in top
of reagent cabinet and sliding reagent cabinet
back.
3. Tighten screws to attach reagent cabinet to main
cabinet.
4.3 WASTE TUBE INSTALLATION
To install the waste tube, proceed as follows:
1. Guide waste tubing through main cabinet into
reagent cabinet and out the side of reagent cabi-
net.
2. Secure plastic waste tube in clamp in lower right of
main cabinet.
See Figure 3-1
.
FIGURE 4-1 MOUNTING MAIN AND
REAGENT CABINETS