a D:GROUP Company
0254EN
– Rev. 06 – 09/2019
Page 34 of 65
10 MONITORING SYSTEM
ALARM OR ERROR MESSAGE
DESCRIPTION
Min. LAF Alarm
Airflow speed in the work chamber under minimum threshold value
Min Pressure Alarm
The pressure of the work chamber is raised under the minimum
alarm value. It can be due to a hole in the gloves or to the opening of
a door.
LAF Triac Failure
Too fast and uncontrollable main motor-fan (LAF)
No encoder input LAF
No input from the fan anemometer installed on the LAF main motor-
fan
No encoder input EXH
No input from the fan anemometer installed in the exhaust duct
Max LAF Alarm
Airflow speed in the work chamber over maximum threshold value
Front glass open
The front window is not closed correctly
Sensors failures
Failure to one of the sensors. Press
“MUTE” key and the
correspondent sensor is displayed
BLACK-OUT
Warning of black-out when the isolator is working
Press “MUTE” to stop the alarm
HEPA Filters Check
Possible clogging of absolute (HEPA) filters
Exhaust Duct Check
Possible clogging of the exhaust duct
Position the window
The front window is not in the correct position
F1 (2, 3, 4, 5) lifetime over
Filter 1 (2, 3, 4, 5) residual lifetime is over
ATTENTION DANGER
Reduced airflow. DO NOT WORK
Warning to pay attention when the speed reduction is activated
(only for k
eyboards with “SPEED REDUCTION” key)
Summary of Contents for GLOVEFAST ASEPTIC
Page 9: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 9 of 65...
Page 26: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 26 of 65 1 2...
Page 46: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 46 of 65...
Page 48: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 48 of 65 15 J DIAGRAM FOR MAINTENANCE OPERATIONS...
Page 49: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 49 of 65 16 FRONTAL DIAGRAM...
Page 50: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 50 of 65 16 A SIDE DIAGRAM...
Page 54: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 54 of 65 18 WIRING DIAGRAM...
Page 55: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 55 of 65...
Page 56: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 56 of 65...
Page 57: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 57 of 65...
Page 58: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 58 of 65...
Page 59: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 59 of 65...
Page 60: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 60 of 65...
Page 61: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 61 of 65...
Page 62: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 62 of 65...
Page 63: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 63 of 65...
Page 64: ...a D GROUP Company 0254EN Rev 06 09 2019 Page 64 of 65...