2. Terminal for outputting recorder and alarm
Attaching terminal
1
)
How to set the terminal
(
1
)
By using cross-slot screwdriver, remove the screws of the
cover. You can remove the cover with wires.
(
2
)
Remove the wires on the cover.
※
Removed cover won't be necessary. However, Please do not
lose it.
(
3
)
Connect the connector for connecting wire to the connector
of terminal for recorder and alarm.
※
Insert the connector in proper direction.
(
4
)
Fix the terminal for outputting recorder and alarm with the
screws that were removed from the main unit.
※
When fixing each terminal to the main unit, do not
tuck the wires.
・
Terminal for outputting recorder
Outputs the data to record the temperature in
the chamber (displayed value) in the recorder.
Output range
DC1
〜
4V
(
Temperature range-20
〜+
80
℃)
DC30mV/
℃、
Accuracy
±
12mV
-22-
・
Terminal for alarm
When the safety feature works, it outputs the unusual
status of the unit at the contact point (b contact point).
Capacity of contact point
AC250V/DC30V Max 3A
Left side
・
・ ・
・
Optional accessory
Cover for attaching terminal for alarm
Optional accessory
Cover for attaching
terminal for outputting
recorder
Terminal
for alarm
Terminal for outputting
recorder