-4-
2-3
Characteristic of temperature rise and fall (reference)
Condition
・
AC-100V 50Hz
・
Center-focused temperature in the chamber
・
No load
Notice
Highest,lowest attainable temperature, accuracy of temperature control and
temperature distribution differ depending on room temperature, heat load, supply
power and voltage.
Attainable chilling temperature of LTI-400E are as follow;
10
℃
while the room temperature is 35
℃
, 4
℃
while the room temperature is 30
℃
.
0
10
20
30
40
50
0
10
20
30
40
50
100
Time(min.)
erature(℃)
60
70
80
90
60
(
1
)
Room temperature: 20
℃
Fan in the chamber:ON
110
120
70
Fan in the chamber: OFF
Fan in the chamber:OFF
Fan in the chamber: ON
0
10
20
30
40
50
0
20
40
60
80
100
200
Time(min.)
ature(℃)
120
140
160
180
60
Room temperature:30℃
220
240
70
Room temperature:35℃
(
2
)
Room temperature: 30 and 35
℃
, fan in the chamber: OFF