-53-
(
4
)
Checking drainage tray
Frost that adheres to refrigeration unit can be melted and
drained out from the chamber by executing forced
defrosting. The drained water will be flowed into the
drainage tray (Approx. 4L) on the back surface and
evaporated. Because of forced defrosting, the tray may
be overflowed with huge quantity of water. Check the tray
whether there's water or not before executing forced
defrosting.
Check the quantity of water in the tray regularly and remove
the cover at the bottom on the left side of the unit to remove
the water with using vessel and etc.
(5
)
Cleaning wire condenser
(
refrigeration unit and natural
air-cooling type condenser
)
Wire condenser (black) is attached on the back surface.
Dirt on the condenser worsens the cooling capability
and cause malfunction on the refrigeration unit.
Check the condition regularly and remove the dirt with
brush and etc.
Wire condenser cover
Before cleaning and caring the product, turn the
mains switch off and disconnect the power plug
from the outlet. Or it may cause electric shock or
damage the product.
※
Above illustration is LTI-400E.
Wire condenser (black) may be hot while the unit
is running or right after the operation.
Also, wire condenser consists of thin wires, which
may hurt you if you touch the edge of
the wire directly.
Do not touch wire condenser while the unit is
running or cleaning and carrying the unit.