4-3
Installation
Do not incline the main unit.
This unit is equipped with refrigeration unit.
So do not incline the unit or carry the unit with
inclining it.
CAUTION
1. Precautions for charting the unit
(
LTI-1200E
)
-19-
Do not hold the wire condenser cover
when carrying or moving the unit. It may
damage the unit.
(1)
Raise the level adjuster (4 points) with using
spanner wrench (subtense: 17) and release the
lock of casters.
(2)
Move the unit to the installation site.
※
Carry the unit when passing through the bumpy
floor so that the casters wont' be damaged.
(3)
At the installation site, adjust the level adjuster
with using spanner wrench (subtense: 17) and
level the unit.
Adjust the level adjuster to fix the casters
and level the unit.
Rotating part
Lowering the unit
Raising the unit
Level adjuster
Caster
(
With 4 level adjusters
)
2. Fixing the product
(
How to use the clamp
)
Clamp is used for fixing the unit on the floor. Port
for inserting anchor bolt is equipped (Diameter :
7mm, 12mm) .
(
1
)
Remove 4 screws (see the right picture) on
the back side of the unit with cross-slot
screwdriver.
(
2
)
Tighten the screws, which were removed
from 2 clamps, on the base by using cross-
slot screwdriver.
(
3
)
Fix the clamp and the floor with 2 anchor
bolts.
※
Anchor bolt is not supplied with the product.
※
Above illustration is LTI-400E.
Clamp
4 screws
Diameter of port: 12mm
Diameter of port: 7mm
2 screws attached to LTI-400E
will be used when using 2-step
double clamp.
Wire condenser
cover (Black)